For Sterile Filtration of Pharmaceuticals and Biological solutions.
Construction
Rigid construction with internationally accepted GMP standard.
Capacity
For 90mm, 142mm, 293mm diameter membrane filters
Material Grade
S.S. 316/S.S. 316L
Finish
Mirror polished to 220 grit
Design
Unipack offers complete systems for aseptically filtering liquid samples. The sterility test in these systems are done by the bacteria-retentive membrane filter method.
Special Features
Bubble Test Arrangement – ‘Bubble Point Test’ is a non-destructive test to check the intergrity of the membrane filter, before and after the filtration. The bubble point is the pressure required to force air through the pores of a wet filter. If a membrane cannot with stand the pressure below bubble point level, it Indicates leak in the filter. A built in system for bubble testing can be provided with the membrane holders by in building a set of liquid & air control valves